Convert simulate cardioversion by activating convert* feature.
Defib discharge- indicator illuminates for two seconds when simulator is shocked by a defib set to 50J, or more.
Connectors PADS only.
Battery powered by one 9 volt alkaline battery. Low battery indicator illuminates when battery needs to be replaced. Battery saver feature powers off simulator automatically when not in use. Battery life is approximately 25 hours.
Three year warranty covers material and workmanship.
Convert
Convert automatically from a running rhythm to a waiting rhythm when simulator is shocked by a defib set to 50J, or more.
Press the convert key to engage. The indicator of the running rhythm will pulse brighter.
Press the key of the rhythm you wish to activate after the defib discharge. This is the waiting rhythm. The indicator of the waiting rhythm will flash on and off.
Shock simulator with defib set to 50J or more. The waiting rhythm becomes the new running rhythm.
To cancel the convert operation before a shock is delivered, press the convert key again, or press the key of the running rhythm. The convert operation cancels automatically after two minutes if a shock is not delivered.